+8615966019173
Zero Point Thermostat
Home - Products - Temperature Measuring Equipment - Zero Point Thermostat - DY-BO-50 Multi-channel Freezing Point Box/Freezing Point Chamber
The DY-BO-50 multi-channel ice point bath/ice point box is a highly stable ice point device that provides a constant 0°C temperature field for the reference junction of thermocouples, thus ensuring the accuracy of temperature measurements. The multi-channel ice point bath/ice point box uses mechanical control to provide a constant 0°C temperature field, and its core function is to maintain a stable 0°C temperature field by freezing an ice-water mixture.
It is primarily used in scientific research, laboratories, and industrial testing fields that require a precise freezing point reference, such as in the calibration of thermocouples, providing a stable and reliable 0°C reference temperature for related measurements and experiments.
Product Advantages
The product structure is meticulously planned and designed to effectively reduce environmental interference on the equipment, ensuring stable operation.
Customizable to customer needs, offering thermocouple types E, J, K, R, S, T, and N.
The thermocouple terminals and compensation wires use high-quality imported components to ensure the accuracy of the equipment's measurements.
The welding points between the thermocouple wires and copper wires in each channel are insulated from each other, ensuring accurate signal transmission.
Equipped with a cooling indicator light for users to easily understand the equipment's operating status.
Powered by a 220V AC power supply, with reliable grounding of the casing to ensure safe use.
Stability: During normal operation, all channels can stably maintain the same value within the tolerance range (to one decimal place), without frequent fluctuations within the tolerance range.
The internally used PT100 sensors have been rigorously calibrated before use.
| Product Name | Multi-channel Ice Point Box/Ice Point Reference | |
| Product Model | DY-BO-24 | DY-BO-50 |
| Channels | 24 Channels | 50 Channels |
| Internal Thermocouple Compensation Wire | Type J Thermocouple Compensation Wire (Other types of compensation wires can be customized) | |
| Stability | Accuracy of all thermocouple channels meets ±0.2℃ Accuracy of RTD channels meets ±0.2℃. | |
| Dimensions | 490mm × 265mm × 340mm | |
| Power Supply | 220V 50HZ | |